Keysight NFA Series Noise Figure Analyzers Service Guide 85
Instrument Messages
Event Messages
–221 Settings Conflict Errors
This is one of the errors in the standard SCPI error range –200 to –299. For all
other errors in this range, see “–200 to –299, Execution Errors” on page 96.
Err# Message Verbose/Correction Information
–221 De-emphasis only
available in FM
The de-emphasis function is only available if FM demod is selected
–221 Function not
available in Zero
Span
The function you are trying to access is not available in zero span
–221 Setting conflict;
<trigger source>
trigger is not
available while input
is <input port>
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not available with the
current input port (RF, IQ, etc.)
–221 Setting conflict;
Differential setting
determined by probe
type
A probe is connected that has a built in Differential setting. The setting
cannot be changed manually
–221 Setting conflict;
Input Z unavailable
when probe sensed
A probe is connected and the Input Z is set based on the probe type. It
cannot be changed manually
–221 Settings conflict; A legal command was received but it could not be executed due to the
current device state
–221 Settings conflict;
*.CSV file format is
not available in this
measurement.
You cannot load or save base instrument traces, as this is not supported by
the Log Plot measurement
–221 Settings conflict; <Q
Param> cannot be
changed when Q same
as I
When the
Q Same as I parameter is set to Yes, the I parameter value is
copied to <Q Param> and the <Q Param> value cannot be changed. Set
Q
Same as I
to No to enable explicit control of the <Q Param> value
–221 Settings conflict; A
Valid User Cal is
required. Optimize
aborted
Optimize Preselector can only be performed if a valid user cal exists and is
applied to current results. Perform a user cal first or apply existing cal
–221 Settings conflict;
Auto Tune not
available in Tracking
Source mode
The Auto Tune feature cannot be used when you are using a Tracking Source
–221 Settings conflict;
BTS gain is not
available in this
Mode
Base Transceiver Station gain correction is not available in some Modes, or in
some measurements (for example, the SA measurement)

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals