EasyManuals Logo

Keysight N8973B User Manual

Keysight N8973B
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #476 background imageLoading...
Page #476 background image
476 Keysight NFA Series Noise Figure Analyzers Service Guide
Post-Repair Procedures
Post-Repair Procedures
Additional Tasks
Calibration Data Backup and Restore
In order for the instrument being repaired to not need all of the instrument
adjustments and performance verification tests to be run after the A5 Disk
Drive is replaced the calibration data will need to be backed up onto an
external drive prior to removing the existing disk drive.
Data Backup
There are two different backup procedures that might need to be followed.
Which backup procedure to be run will depend on whether the software will
run or not.
If the Instrument Software Application Will Run
1. Connect a USB keyboard and mouse to two of the instrument rear panel
USB ports
2. Connect a USB FLASH drive to one of the front panel USB port.
3. Press System, Alignments, Backup or Restore Align Data.
4. After the XSA application closes the Alignment Data Wizard window as
shown in Figure 16-1 will appear.
While the backing up and restoring of the calibration data to the new disk
drive will return the instrument performance to the state that it was prior
to a repair this does not guarantee that all instrument performance
parameters meet their specified values. A full instrument calibration
would be required to verify this.

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals