Keysight NFA Series Noise Figure Analyzers Service Guide 423
Assembly Replacement Procedures
RF Front End Assembly (N8976B)
Replacement
1. Refer to Figure 15-25. Place the RF Front End Assembly into the chassis.
Replace the four screws (2). Torque to 9 inch-pounds.
2. Reattach ribbon cable W17 to the RF Front End Assembly.
3. Reattach the cables W3, W4, W5, W7, W15 (or W36), and W20 to the RF
Front End Assembly (1). Torque the semi-rigid cables to 10 inch-pounds.
4. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
5. Replace the instrument top brace. Refer to the Top Brace replacement
procedure.
6. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.