Keysight NFA Series Noise Figure Analyzers Service Guide 263
Analog/Digital IF Troubleshooting
25 MHz BW IF Section
A3 Digital I.F. Troubleshooting
Verifying the 22.5 MHz Input Power
1. Perform an instrument shut down.
2. Remove the cover of the analyzer. Refer to Chapter 15, “Assembly
Replacement Procedures”, on page 389 in this service guide.
3. Turn the instrument over so that the bottom is facing up.
4. Turn on the analyzer and wait for the instrument to complete the boot up
process.
5. Press System, Alignments, Auto Align, Off
6. Press Input, RF Calibrator, 50 MHz
7. Verify the 50 MHz signal is at 25 dBm by pressing MODE/MEAS,
Spectrum Analyzer, OK, FREQ, 50 MHz, Span, 1MHz, Peak Search on the
analyzer. The marker readout should be 50 MHz at 25 dBm ± 3 dBm. If
this reference signal is measuring incorrectly, see Chapter 4, “RF Section
Troubleshooting (N8973B, 74B, 75B Analyzers)”, on page 133 in this
service guide.
8. To continue verifying press FREQ, Zero Span. Verify the input attenuator
on the analyzer is set to 10 dB. (Look near the top of the display near the
center and verify Atten: 10 dB)
If the analyzer is not in 10 dB of input attenuation press AMPTD,
Attenuation, Mech Atten 10 dB.

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals