96 Keysight NFA Series Noise Figure Analyzers Service Guide
Instrument Messages
Event Messages
–200 to –299, Execution Errors
This section lists all messages in the range except –221 error messages. For
details of –221 messages, see “–221 Settings Conflict Errors” on page 85.
Note that Execution Errors are divided into subclasses, as follows:
–221 Settings
conflict;Function not
available while
measurement is
running
Settings change is not allowed while measurement is running. You must stop
the current measurement if you wish to change the settings
–221 Settings
conflict;function
unavailable with this
EMC Standard
–221 Settings conflict;Log
Percent Auto Step
Rule does not apply
in Swept Scan Type.
The Log Percent rule is not available when Scan Type=Swept Scan because
we are always doing linear sweep
–221 Settings conflict;QPD
+ EMI Average + RMS
Average is not
allowed
You cannot turn on all 3 EMI detectors together. You must turn off one of the
EMI Detectors before you turn this on
–221 Settings
conflict;Range <0> is
turned off as total
range points > 40001
Max of Total range points is 400000. Reduce Scan Points or increase Step
Size in order to turn on that range
–221 Settings conflict;RF
Input 2 is not
available above 1GHz
If the frequency range is set above 1GHz, you cannot change to RF Input 2
–221 Settings
conflict;Scan Time &
Points do not apply
in current Step/Time
Control.
The Scan Time & Points are not available when Step/Time Control is set to
Step & Dwell
–221 Settings
conflict;Step Size &
Meas Time do not
apply in current
Step/Time Control.
The Step Size & Meas Time are not available when Step/Time Control is set to
Scan Time&Pts
Err# Message Verbose/Correction Information
–21x Trigger errors
–22x Parameter error

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals