Keysight NFA Series Noise Figure Analyzers Service Guide 97
Instrument Messages
Event Messages
–23x Data corrupt or stale (invalid data)
–24x Hardware error
–25x Mass storage error
–26x Expression data error
–27x Macro error
–28x Program error (a downloaded program-related execution error)
–29x Memory use error
Err# Message Verbose/Correction Information
–200 All ranges are off.
Turn on at least a
range
There are no range turn on in scan table. You need to turn on at least a range to
initiate a scan
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom
–200 Execution Error A program execution error has occurred. The exact problem cannot be
specifically identified
–200 Execution error;
Carrier frequency
outside device’s
transmit band
The entered channel/carrier frequency is not within the range of your current
mode setup selection of standard and device
–200 Execution error;
Invalid GSM burst
timing
A GSM-like burst was acquired, but its timing is not valid. Ensure the correct
Burst Type has been selected
–200 Execution error;
Invalid IP address
The IP address supplied is either not valid or does not belong to a compatible
Signal Generator. Please check the IP address and instrument connection and
try again
–200 Execution error;
Invalid Marker Trace.
Cannot place markers on the reference trace, because the reference trace is
currently turned off or has no data
–200 Execution Error; No
peak found.
No signal peak was found within the defined parameters of the search. (Note:
for ESA/PSA compatibility, the Swept SA measurement uses 780 for this error
number; all other measurements use –220)
–200 Execution error; No
ranges are defined.
Activate a range.
There are no active ranges in the range table. You will need to activate at least
one range
–200 Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by the signal
peak being too low in amplitude, or it could be due to excessive CW input
signal, alignment error, or hardware failure