Keysight NFA Series Noise Figure Analyzers Service Guide 95
Instrument Messages
Event Messages
–221 Settings conflict;
Trigger is not
available with span >
0 Hz.
–221 Settings conflict; Tx
Band Spur meas does
not support this
frequency band.
The transmit band spur measurement does not support all of the
commercially available frequency bands. You need to change your selection
under
Mode Setup, Radio, Band to one of the supported bands
–221 Settings conflict; Tx
Band Spur measurement
is not defined for
mobiles.
Only base station testing is available
–221 Settings
conflict;Administrat
or privileges
required
You must be logged in with administrator privileges to do this. Log out and
log back in as the Administrator, then restart the SA application
–221 Settings
conflict;Auto Scan
Time/Meas Time do not
apply in Stepped Scan
Type.
The Auto Scan Time/Meas Time are not available when Scan Type =Stepped
Scan
–221 Settings
conflict;Auto Scan
Time/Meas Time do not
apply in Stepped Scan
Type.
The Auto Scan Time/Meas Time are not available when Scan Type =Stepped
Scan
–221 Settings
conflict;ClearList &
Start not available
when ScanSeq =
(Re)measure
Cannot perform Clear List & Start during (Re)measure, because we require
the frequency information of the peaks in signal list to perform (Re)measure
–221 Settings conflict;EMI
Detectors and Average
detector can’t be
used together
You cannot turn on any EMI Detector and Average detector together. They
are always mutually exclusive
–221 Settings
conflict;Freq > 1 GHz
is not available
while RF Input 2
enabled
Frequency is limited to 1GHz while RF Input 2 is enabled
Err# Message Verbose/Correction Information