Keysight NFA Series Noise Figure Analyzers Service Guide 103
Instrument Messages
Event Messages
–250 Mass storage error;
Can only import
single trace .csv
files
Trace files containing multiple traces can not be imported. However, if you
need to recall multiple traces you can use the
Save and Recall functions,
rather than the Import and Export functions
–250 Mass storage error;
Can only load an
Antenna Unit into
Correction 1
The only Correction register that supports Antenna Units is number 1. You have
attempted to load an Ampcor file that contains antenna units into another
register
–250 Mass storage error;
Cannot make
The directory or file cannot be created
–250 Mass storage error;
Different Antenna
Unit already in use
Attempt to import Corrections file with Antenna Unit that differs from an in-use
correction
–250 Mass storage error;
Directory not found
The system cannot find the path specified
–250 Mass storage error;
Failed to Load trace.
Bad file format.
The load trace operation could not be completed, as the input file was not in
the expected format. You can only load traces that were previously saved using
the
Save, Trace feature
–250 Mass storage error;
File <filename> wrong
type
Attempt to import a data file that is not the proper type for this operation
–250 Mass storage error;
File <filename> and
instrument version
mismatch
While opening a file, there was a mismatch between file version or model
number with instrument version or model number. The import still tried to load
as much as possible, but you should check it closely
–250 Mass storage error;
File contains
incorrect data for
this operation
There is a mismatch between the file data type of the file specified and the
destination indicated. For example, a correction set cannot be loaded/imported
into a limit line
–250 Mass storage error;
File empty
Cannot save trace because it contains no data. Check that the trace is turned
on and contains some valid data
–250 Mass storage error;
Invalid register
number for *SAV or
*RCL Mass Storage
error
You have used the *SAV command to save a state to a non existent state
register.
Or
You have used the *RCL command to recall a state register that wasn’t
previously saved with the *SAV command
–250 Mass storage error;
Lock violation
The process cannot access the file because another process has locked a
portion of the file
Err# Message Verbose/Correction Information