180   Keysight NFA Series Noise Figure Analyzers Service Guide
RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
— Turn the instrument on and allow it to complete its full boot up process to 
the signal analyzer application.
— Turn Auto Align off by pressing System, Alignments, Auto Align, Auto 
Align, Off on the analyzer.
— Verify all power supply LED's on the Front End Troubleshooting board are 
turned on. Under the J2 connector, the supplies are +15 VDC, +9 VDC, 
+5.2 VDC, and −15 VDC. Under the J1 connector, the supplies are +15 VDC, 
−15 VDC, +9 VDC and +5.2 VDC. Verify both input attenuator power 
supplies. DS9 under J3 should be on if the +25V supply is up for Attenuator 
A. DS10 under J4 should be on if the +25V supply is up for Attenuator B. If 
any of the power supply LED's are not turned on, see Chapter 2, “Boot Up 
and Initialization Troubleshooting”.
—Press AMPTD, Attenuation, Mech Atten, 0 dB
4. Input Attenuator A Control Logic Verification 
All the attenuation LED's on the Front End Troubleshooting board should be off 
except for DS9 and DS10, +25V supply for Attenuator A and Attenuator B. 
Resistors on the board can get very hot.
The analyzer will display an error dialogue box when booting up. Simply 
click “OK” so the instrument can complete its boot process. Additionally, 
the analyzer noise floor will be near the top of the display since the RF 
assemblies are not powered up. This is normal with the Front End 
Troubleshooting board connected to the A15 Front End Driver Assembly