484 Keysight NFA Series Noise Figure Analyzers Service Guide
Post-Repair Procedures
Post-Repair Procedures
FPGA Synchronization
There is FPGA (Field-Programmable Gate Array) program code on many 
different assemblies inside of the instrument, and all of these must be of a 
compatible version with the instrument software revision. In order to ensure 
that this requirement is met when either an assembly is replaced that contains 
FPGA code on it, or if the A5 Disk Drive is replaced with a drive that has a 
different version of software than the previous drive, you will need to take 
certain steps to ensure that the instrument operates properly.
Updating the Instrument FPGA Code
All instruments have a utility included in them that will update the FPGA code 
on all of the different assemblies in the instrument that have it. When this 
utility is run it will detect the assemblies that need to be updated and will 
update them to the correct revision.
1. Close the instrument application software by pressing System, Exit 
Program, OK using the front panel keys.
2. Connect a USB mouse to one of the instrument front panel USB ports.
3. Using the mouse double-click on the Windows Explorer link on the 
desktop and navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
4. In this folder find and execute the file named: 
FPGA_Prog.bat
Even if an instrument appears to work properly when an assembly with 
FPGA code is replaced, that does not mean that the FPGA code is 
completely compatible with the software version installed in the 
instrument. There could be incompatibilities that will only be seen under 
certain measurement conditions.
Once you start the FPGA programming process you MUST NOT interrupt 
the process for any reason. This would include turning the instrument off 
or unplugging the power cord to the instrument. Doing so will result in an 
inoperative instrument, requiring the affected assembly to be replaced.