PLZ-U Remote Control 6-27
6.5.7 Input State Commands
INP/OUTP
Turn ON/OFF the load. Or, queries the current setting.
When the loads on multiple channels are turned on simultaneously, there may be an
offset of about 10 ms for the loads on different channels to turn on.
Set command: INPut[:STATe][:IMMediate]
{OFF|ON|0|1}
OUTPut[:STATe][:IMMediate]
{OFF|ON|0|1}
Query command: INPut[:STATe][:IMMediate]?
OUTPut[:STATe][:IMMediate]?
• Program data
Value OFF(0) Turn the load off
ON(1) Turn the load on
INP:EXT:LPOL/OUTP:EXT:LPOL
Sets the polarity of the channel control signal used to turn the load on. Or, queries
the current setting.
Set command: INPut:EXTernal:LPOLarity
{POSitive|NEGative}
OUTPut:EXTernal:LPOLarity
{POSitive|NEGative}
Query command: INPut:EXTernal:LPOLarity?
OUTPut:EXTernal:LPOLarity?
• Program data
Value POSitive High active
NEGative Low active
INP:EXT:REF/OUTP:EXT:REF
Sets the external voltage reference of the load unit. Or, queries the current setting.
Set command: INPut:EXTernal:REFerence
{NONE|MAIN|SUB}
OUTPut:EXTernal:REFerence
{NONE|MAIN|SUB}
Query command: INPut:EXTernal:REFerence?
OUTPut:EXTernal:REFerence?
• Program data
Value NONE No external control
MAIN External input
CC control for CC and CC+CV modes
CR control for CR and CR+CV modes
CV control for CV mode
SUB External input
CV control for CC+CV and CR+CV modes