6-28 Remote Control PLZ-U
INP:PROT:CLE/OUTP:PROT:CLE
Clears the alarm.
Set command: INPut:PROTection:CLEar
OUTPut:PROTection:CLEar
INP:TIM/OUTP:TIM
Sets the cutoff time. Or, queries the current setting.
Set command: INPut[:STATe]:TIMer[:STATe]
{<value>|MINimum|MAXimum}
OUTPut[:STATe]:TIMer[:STATe]
{<value>|MINimum|MAXimum}
Query command: INPut[:STATe]:TIMer[:STATe]?
OUTPut[:STATe]:TIMer[:STATe]?
• Program data
Value 1 to 99999 or 0 (function off)
Resolution 1
Unit S
INP:DEL/OUTP:DEL
Sets the delay time until the load is turned on. Or, queries the current setting.
Set command: INPut[:STATe]:DELay
{<value>|MINimum|MAXimum}
OUTPut[:STATe]:DELay
{<value>|MINimum|MAXimum}
Query command: INPut[:STATe]:DELay?
OUTPut[:STATe]:DELay?
• Program data
Value 0.001 to 1.000 or 0 (function off)
Resolution 0.001
Unit S