PLZ-U Remote Control 6-29
Sequence commands
Programs a series of steps. For details, see section 5.6, “Sequence Function.
If a command includes notations such as {<conductance>|<cur-
rent>|<power>|<voltage> the program data corresponding to the selected
program mode is selected.
Creating a sequence
PROG:MEMO
Sets the program memo (comment). Or, queries the current setting.
Up to 11 characters can be specied.
Set command: PROGram[:SELected]:MEMO "<string>"
Query command: PROGram[:SELected]:MEMO?
PROG:LOOP
Sets the number of program loops. Or, queries the current setting.
Set command: PROGram[:SELected]:LOOP <value>
Query command: PROGram[:SELected]:LOOP?
Program data
Value 1 to 9998 or 9999 (innite loop)
Resolution 1
PROG:LINP/PROG:LOUT
Sets the load on/off condition after the sequence ends. Or, queries the current set-
ting.
Set command: PROGram[:SELected]:LINPut
{OFF|ON|0|1}
PROGram[:SELected]:LOUTput
{OFF|ON|0|1}
Query command: PROGram[:SELected]:LINPut?
PROGram[:SELected]:LOUTput?
Program data
Value OFF(0) Turn the load off
ON(1) Turn the load on
PROG:LVAL
Sets the current value after the end of the program. Or, queries the current setting.
Set command: PROGram[:SELected]:LVALue
{<conductance>|<current>}
Query command: PROGram[:SELected]:LVALue?
Returns the current value in response to PROG:LVAL?.
Example) When a current value of 9.876 A is specied
Returns 9.876.
NOTE

Table of Contents

Other manuals for Kikusui PLZ-U Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kikusui PLZ-U Series and is the answer not in the manual?

Kikusui PLZ-U Series Specifications

General IconGeneral
BrandKikusui
ModelPLZ-U Series
CategoryControl Unit
LanguageEnglish

Related product manuals