Important Information
9287C_002-462e-03.11 Page 5
2. Important Information
Please keep to the following rules without fail. This will en-
sure your personal safety when working, and assure long,
trouble-free performance by the instrument.
2.1 For your Safety
This instrument has been tested thoroughly and it left
the works in a perfectly safe condition. To maintain this
condition and assure safe operation, the user must ob-
serve the directives and warnings contained in these in-
structions
The force plate must be installed, operated and main-
tained only by persons who are familiar with it and
adequately qualified for their particular tasks
When it must be assumed that safe operation is no
longer possible, the instrument must be taken out of
operation and secured against unintentional use
It must be assumed that safe operation is no longer
possible if:
the instrument is visibly damaged
it no longer functions
it has been in lengthy storage under adverse con-
ditions
it has received rough treatment during transport
According to safety requirements the following has to be
observed:
The force plate should be mounted as written in chap-
ter 4.
The force plate has to be connected to earth using the
Potential Equalization Conductor provided in the scope
of delivery to meet the ä-standards regarding Earth
Leakage Current.
The data acquisition computer and persons who are in
direct contact with it have to remain in a distance of
1,5 m from the patient at all times for the same reason.