Multicomponent Force Plate for Biomechanics Type 9287C…
2.2 ä-Conformity
The Force Plate Type 9287C… is conforming to ä direc-
tives according to the Declaration of Conformity enclosed
with this instruction manual.
2.3 How to Treat the Instrument
The force plate may be used only under the specified
environmental and operating conditions
Protect the signal output against dirt and do not touch
it with your fingers. When the connection is not being
used, cover it with the cap provided
When the force plate is not in use, protect it by keeping
it in the packing case supplied
When performing long-time measurements, make sure
that the temperature of the force plate remains as con-
stant as possible
2.3.1 With External Amplifier (Type 9287C)
The insulation resistance is crucially important with pie-
zoelectric measurements. It must be around 10
14
Ω (but
at least 10
13
Ω)
To obtain this resistance, all plug and socket connec-
tions must be kept meticulously clean and dry
The insulation resistance can be measured with the in-
sulation tester Type 5493
The connecting cable from force plate to charge ampli-
fier is highly insulating. Use only the proper cable
Do not remove the connecting cable from the force
plate
Pag
e 6 9287C_002-462e-03.11