EasyManua.ls Logo

Konica Minolta CA-410 - Specifications of Each Probe

Konica Minolta CA-410
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
Specifications of Each Probe
CA-P427 CA-P427H
Measurement area Ø27 mm Ø27 mm
Acceptance angle ±2.5° ±2.5°
Accuracy guaranteed measurement distance 30±10 mm 30±10 mm
Display range Luminance 0.0001 to 5000 cd/m
2
0.0001 to 30000 cd/m
2
Chromaticity Displayed in 4 digits Displayed in 4 digits
Luminance Accuracy guaranteed luminance range 0.001 to 5000 cd/m
2
0.01 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.001 - cd/m
2
±9% ---
0.01 - cd/m
2
±2% ±9%
0.1 - cd/m
2
±1.5% ±2%
1 - cd/m
2
±1.5% ±1.5%
10 - cd/m
2
±1.5% ±1.5%
100 - cd/m
2
±1.5% ±1.5%
Repeatability
(2σ)
*1
AUTO 0.001 - cd/m
2
10% ---
0.01 - cd/m
2
1% 10%
0.1 - cd/m
2
0.40% 1%
1 - cd/m
2
0.10% 0.40%
10 - cd/m
2
0.10% 0.10%
100 - cd/m
2
0.10% 0.10%
Chromaticity Accuracy guaranteed luminance range 0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.01 - cd/m
2
±0.003 ---
0.1 - cd/m
2
±0.002 ±0.003
1 - cd/m
2
±0.002 ±0.002
10 - cd/m
2
±0.002 ±0.002
100 - cd/m
2
±0.002 ±0.002
In 100 cd/m
2
(for monochrome)
*2
100 cd/m
2
±0.003 ±0.003
Repeatability
(2σ)
*1
AUTO 0.01 - cd/m
2
0.0035 ---
0.1 - cd/m
2
0.0015 0.0035
1 - cd/m
2
0.0004 0.0015
10 - cd/m
2
0.0003 0.0004
100 - cd/m
2
0.0002 0.0003
Flicker (CA-310 Mode)
*6
Flicker
(Contrast)
Measurement luminance range 5 to 1500 cd/m
2
30 to 9000 cd/m
2
Range to be measured (flicker frequency) 0.25 to 65 Hz 0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave
± 0.4% ± 0.4%
60 Hz, AC/DC 10% sine wave
± 0.7% ± 0.7%
Repeatability (2σ)
20 to 65 Hz,
AC/DC 10% sine wave
0.3% 0.3%
Flicker
(JEITA)
Measurement luminance range 5 to 1500 cd/m
2
30 to 9000 cd/m
2
Range to be measured (flicker frequency) 0.42 to 65 Hz 0.42 to 65 Hz
Accuracy
30 Hz, AC/DC 4% sine wave
±0.35 dB ±0.35 dB
30 Hz, AC/DC 1.2% sine wave
±0.35 dB ---
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
0.1 dB 0.1 dB
30 Hz, AC/DC 1.2% sine wave
0.3 dB ---
Specifications of Each Probe (1)

Table of Contents