EasyManua.ls Logo

Konica Minolta CA-410 - About User Calibration

Konica Minolta CA-410
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
83
About User Calibration
User calibration allows users to set their own correction coecients to calibration channels of the instrument by setting
calibration values (Lv x y, Lv u’v’, or X Y Z) on the instrument. Afterward, the value corrected with the specified correction
coecients will be displayed/output for every measurement.
User calibration can be performed to apply the following corrections to measurements based on the specified correction
coecients.
(1) Correction of variation of readings that occur due to the deviation of spectral sensitivity from the CIE 1931 color-
matching function
(2) Correction of dierence in readings between instruments when two or more instruments are used
Single-point calibration or RGB+W matrix calibration can be selected for user calibration. User calibration ensures accurate
measurement of luminance and chromaticity near the calibration light source.
• When obtaining the calibration value by measuring the calibration light source with a measurement reference device, be
sure to set the reference device and the instrument at the same position and angle from the calibration light source and
measure the same measuring area. The instrument may not be calibrated properly if the measurement conditions are
dierent from those of the reference device.
• During measurement, use a constant-voltage power supply for illuminating the calibration light source to ensure
conditions are as stable as possible.
User calibration can be performed for every calibration channel except for CH00.
User calibration can be performed through the following methods.
(1) By Measurement (Single-point calibration) .................................P. 84
(2) By Measurement (RGB+W matrix calibration) ............................. P. 86
(3) Selecting from Saved Data .......................................................... P. 88
(4) By inputting edit correction coecients .....................................P. 90
User calibration cannot be performed for calibration channel CH00.
(CH00 is a calibration channel for performing measurement in accordance with Konica Minoltas calibration standard.)
The user calibration correction coecients are used in common for each color mode (L
v
x y, L
v
u’v’, L
v
Tcp duv, X Y Z, and
Dominant wavelength).
CA-DP40
Calibration/
Settings

Table of Contents

Related product manuals