EasyManua.ls Logo

Konica Minolta CA-410 - Zero Calibration; Performing Zero Calibration

Konica Minolta CA-410
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66
Zero calibration
Zero calibration performs zero point adjustment while blocking entry of light into the probes receptor.
A shutter located inside the probe automatically closes to block light during zero calibration of the instrument. Zero
calibration must be performed whenever the POWER switch is set to ON ( | ).
1. Performing Zero Calibration
Notes on zero calibration
• Perform zero calibration if the ambient temperature has changed. If “Auto 0-cal” is set to ON, zero calibration is
performed automatically when the ambient temperature has changed. (Refer to [Option] - [0-Cal])
• Perform zero calibration at any time by pressing the [0-Cal] key.
• Just in case, do not direct the end of the probe toward an illuminant with high luminance during zero calibration.
• Never press any keys during zero calibration. Doing so will cause completion of zero calibration to take more time.
• Zero calibration will be performed simultaneously with all the connected probes.
Operating Procedure
Measuring probe Data processor
(1)
(2)
Before starting zero calibration, check that a probe is connected to a probe connector.
1
Check that the POWER switch is set to
ON ( | ).
2
Press the [0-CAL] key.
Measurement will start automatically at the end of
zero calibration.

Table of Contents