94
Deleting Calibration Data
Follow the procedure below to delete saved values.
Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key.
The setting screen is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[User Calib], and then press the [RIGHT]
or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
MeasCond
UserCalib
Target
Option
System
MeasureandRegister
CopyandRegister
EditCorrectionCoef
Set ID
DeleteCalibData
3
Press the [UP] or [DOWN] key to select
[Delete Calib Data], and then press the
[RIGHT] or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
MeasCond
UserCalib
Target
Option
System
MeasureandRegister
CopyandRegister
EditCorrectionCoef
Set ID
DeleteCalibData
4
The screen for selecting a probe is
displayed. Press the [UP] or [DOWN] key
to select the probe to register, and then
press the [RIGHT] or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
MeasCond
UserCalib
Target
Option
System
10
09
08
07
06
05
04
03
02
01
MeasureandRegister
CopyandRegister
EditCorrectionCoef
Set ID
DeleteCalibData
5
The screen for selecting data to be
deleted is displayed. Press the [UP] or
[DOWN] key to select [All CH] or [Select
CH], and then press the [RIGHT] or
[ENTER] key.
User Calib > Delete
MENU
All CH
Select CH

Table of Contents