32
CA-P410 CA-P410H
Measurement area Ø10 mm Ø10 mm
Acceptance angle ±5° ±5°
Accuracy guaranteed measurement distance 30±5 mm 30±5 mm
Display range Luminance 0.0001 to 5000 cd/m
2
0.0001 to 30000 cd/m
2
Chromaticity Displayed in 4 digits Displayed in 4 digits
Luminance Accuracy guaranteed luminance range 0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.001 - cd/m
2
--- ---
0.01 - cd/m
2
±2.5% ---
0.1 - cd/m
2
±2% ±2.5%
1 - cd/m
2
±2% ±2%
10 - cd/m
2
±1.5% ±2%
100 - cd/m
2
±1.5% ±1.5%
Repeatability
(2σ)
*1
AUTO 0.001 - cd/m
2
--- ---
0.01 - cd/m
2
2% ---
0.1 - cd/m
2
0.60% 2%
1 - cd/m
2
0.20% 0.60%
10 - cd/m
2
0.10% 0.20%
100 - cd/m
2
0.10% 0.10%
Chromaticity Accuracy guaranteed luminance range 0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.01 - cd/m
2
±0.006 ---
0.1 - cd/m
2
±0.002 ±0.006
1 - cd/m
2
±0.002 ±0.002
10 - cd/m
2
±0.002 ±0.002
100 - cd/m
2
±0.002 ±0.002
In 100 cd/m
2
(for monochrome)
*2
100 cd/m
2
±0.003 ±0.003
Repeatability
(2σ)
*1
AUTO 0.01 - cd/m
2
0.0070 ---
0.1 - cd/m
2
0.0020 0.0070
1 - cd/m
2
0.0008 0.0020
10 - cd/m
2
0.0005 0.0008
100 - cd/m
2
0.0003 0.0005
Flicker (CA-310 Mode)
*6
Flicker
(Contrast)
Measurement luminance range 15 to 3000 cd/m
2
or more 90 to 18000 cd/m
2
or more
Range to be measured (flicker frequency) 0.25 to 65 Hz 0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave
± 0.4% ± 0.4%
60 Hz, AC/DC 10% sine wave
± 0.7% ± 0.7%
Repeatability (2σ)
20 to 65 Hz,
AC/DC 10% sine wave
0.3% 0.3%
Flicker
(JEITA)
Measurement luminance range 15 to 3000 cd/m
2
or more 90 to 18000 cd/m
2
or more
Range to be measured (flicker frequency) 0.42 to 65 Hz 0.42 to 65 Hz
Accuracy
30 Hz, AC/DC 4% sine wave
±0.35 dB ±0.35 dB
30 Hz, AC/DC 1.2% sine wave
±0.35 dB ---
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
0.1 dB 0.1 dB
30 Hz, AC/DC 1.2% sine wave
0.3 dB ---
Specifications of Each Probe (2)