74
Setting the Interval Measurement
7
Press the [UP] or [DOWN] key and the
[RIGHT]/[LEFT] keys to set a value, move
to [Complete], and then press the
[ENTER] key.
* The interval time can be set to between 10 to
3,600 sec and the interval number to between 1
to 9,999. If the entered value is out of the
available range, an alert screen is displayed. Press
the [ENTER] key to return to the previous screen
and reset the value.
Press the [ESC] key to cancel configuration of the
setting and restore the original value.
Input Error
The input data is incorrect
Input correct value
OK
8
Press the [ESC] key twice.
The interval measurement screen (Hold) is displayed
on the LCD screen. Press the [HOLD] key to begin
interval measurement.
The interval measurement setting and value will be
kept even if the POWER switch is set to OFF ().