6
2. Before Use
3.
Select the language. (Appears only when turned O for the rst
time)
The Language Selection screen is displayed. Select the language to use.
4.
Press the [OK] to select the language.
The language can be switched at any time. (
¨
P119)
5.
Dark calibration.
The CL-7F measuring system must be calibrated before use. Turn Light Selection
Ring to calibration indication. "Dark calibration in progress. Please wait" and the
status bar will appear while calibrating. The Display Mode Selection screen will
appear when operational.
NOTE
Dark calibration is performed when new batteries are used, 24 hours passed since last use
or there is a big change in temperature between turning power OFF and ON.
Except the cases above, dark calibration after power ON is skipped.