EasyManuals Logo

Konica Minolta CL-70F User Manual

Konica Minolta CL-70F
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
6
2. Before Use
3.
Select the language. (Appears only when turned O for the rst
time)
The Language Selection screen is displayed. Select the language to use.
4.
Press the [OK] to select the language.
The language can be switched at any time. (
¨
P119)
5.
Dark calibration.
The CL-7F measuring system must be calibrated before use. Turn Light Selection
Ring to calibration indication. "Dark calibration in progress. Please wait" and the
status bar will appear while calibrating. The Display Mode Selection screen will
appear when operational.
NOTE
Dark calibration is performed when new batteries are used, 24 hours passed since last use
or there is a big change in temperature between turning power OFF and ON.
Except the cases above, dark calibration after power ON is skipped.
A blue highlight
will appear
behind the
selected button.
Language Selection
Screen
Language Selection
Conrmation Screen
Dark Calibration
Process Screen

Table of Contents

Other manuals for Konica Minolta CL-70F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CL-70F and is the answer not in the manual?

Konica Minolta CL-70F Specifications

General IconGeneral
BrandKonica Minolta
ModelCL-70F
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals