68
5. Measuring Light Sources [Measurement Screen]
CAUTION
Do not trigger ash while skin or other objects are in contact with the ash
tube. Do not touch the ash tube after repeated ashes. (It may cause
burns.)
Do not trigger ash while near the eyes of people or animals. (It may
temporarily affect vision.)
The ash may be triggered suddenly. Because there is the possibility of
burns or negative effects on vision, please handle with care.
NOTICE
If the ash output power is too weak compared to the surrounding light, the
meter, when set in cordless mode, may not detect the ash output. In this case,
use "5-4 Measurement in Cord (PC) Flash Mode". (
¨
P69)
Pulsed light sources such as uorescent lights or special lighting could cause
the meter to take cordless ash measurements in rare cases. In these situations,
use "5-4 Measurement in Cord (PC) Flash Mode". (
¨
P69)
If the Light Receptor
1
detects a sudden and bright change in lighting
intensity, the meter may mistakenly take a measurement. To avoid this, take "5-4
Measurement in Cord (PC) Flash Mode". (
¨
P69)
Do not use Cordless Flash Mode for sources which gradually increase to peak
power, such as traditional ash bulbs, as they cannot be detected and measured.
Be sure to take "5-4 Measurement in Cord (PC) Flash Mode". (
¨
P69)
When using Cordless Flash Mode, the LCD screen illumination dims during
measurement, and the LCD is illuminated for 3 seconds after measurement. To
cancel standby mode, touch the screen or press the Menu Button
6
.
In the Cordless Flash mode, the measured color temperature may change
depending on the shutter speed set for ash measurement and the amount of
ambient light present when the measurement is made.
NOTE
When using the meter in Cordless Flash Mode, it is possible to mount the meter to a
light stand, tripod or similar support using the Tripod Socet
0
.
[Over] or [Under] will be displayed if the light source illumination is too bright or not
bright enough, or if the color temperature is out of the measurement range, when the
Measuring utton is pressed. When the color temperature is outside the calculation
range, --- is displayed. In this case, adust the brightness or color temperature of light
source, or switch the light range. 
¨
P78)