Chapter - 5 Initial Startup / 5.2 Preparing the Instrument after Installation
21
Connecting the
Temperature Sensor
If desired, connect the temperature sensor.
Calibrating the
Temperature Sensor
If a temperature sensor is connected, the sensor must
first be calibrated before operating the instrument.
The calibration consists of two temperature
measurements:
Temperature measurement with the connected
temperature sensor
Temperature measurement with a specified
thermometer.
The P3000 determines the temperature difference
between the sensor and the specified thermometer and
corrects the deviation of the sensor for all following
measurements.
Performing the calibration: see Chapter 7.4 Calibrating the
Temperature Sensor.