Chapter - 7 Settings / 7.4 Calibrating the Temperature Sensor
51
7.4 Calibrating the Temperature Sensor
The calibration consists of two temperature measurements:
Temperature measurement with the temperature
sensor
Temperature measurement with a specified
thermometer.
The temperatue difference between the sensor and the specified thermometer is
determined and compensated.
The temperature sensor must be calibrated:
before the first use of the sensor
once a year
if necessary, i.e. when result accuracy lies outside of
the expected range.
Prerequisites For calibration you need:
a temperature sensor
a certified thermometer
a vessel with water
Calling the Function
Open the settings menu with .
Select
Calibration
with the buttons
and
and
press
.
The dialog window Calibration opens.