EasyManua.ls Logo

kruss P3000 - Calibrating the Measurement Range

kruss P3000
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter - 7 Settings / 7.3 Calibrating the Measurement Range
44
7.3 Calibrating the Measurement Range
The calibration consists of two measurements:
Blank measurement of the empty sample chamber.
Quartz standard measurement.
For the quartz standard measurement the optical rotation of the quartz standard at 20
°C and the actual temperature of the quartz are entered. Using this data the polarimeter
automatically determines the nominal value for the optical rotation at the prevailing
actual temperature.
Calibration should be performed in regular intervals. Calibration should be performed
immediately, if result accuracy fails to lie in the desired range.
Opening the Menu
Open the settings menu with .
Select
Calibration
with the buttons
and
.
Open the dialog window Calibration
with
.

Table of Contents

Related product manuals