Chapter - 6 Routine Operation / 6.2 Executing a Measurement
29
6.2 Executing a Measurement
After changing the polarimeter tube, changing of the dilution medium or after opening
the polarimeter tube, a blank measurement of the polarimeter tube must be performed
first.
Determining the Blank
Open the system setting with .
Select Blank with the buttons and .
Insert the empty polarimeter tube into the sample
chamber.
Close the lid of the sample chamber.
Press Start.
Accept the measured blank with Save.
The blank is saved and is subtracted from all following
measurement results.
For a detailled description see Chapter 7.5 Blank.