Chapter - 7 Settings / 7.4 Calibrating the Temperature Sensor
54
Accept the nominal temperature with .
Nominal
temperature
Input of the value measured with the
specified thermomter.
Actual
temperature
Value measured with the temperature
sensor.
Press Ok.
The temperature difference is calculated and all
following temperatures measured with the sensor are
corrected with this value. The program returns to the
main menu.