22
6. Assembly
1
2
Fig. 14 Lamp housing 107/2, opened
1 Mount with halogen lamp
2 Collector
Caution!
Do not remove the new lamp’s dust cover
until you have installed the lamp. Avoid
fingerprints on the lamp.
• Insert the new 12V 100W lamp (14.1) with the
dust cover straight into the socket until it
stops. Be sure that the lamp is inserted
straight.
• Remove the lamp’s dust cover.
• Replace the housing and fasten it in place us-
ing the fastening screw.
• Place the lamp housing in the transmitted
light lamp housing receptacle (fig. 15) and fas-
ten it with the clamping screw on the side.
Abb. 15 Leica DM2500
1 Lamp housing for transmitted light
1