EasyManua.ls Logo

Leica DM2000 - Start-up: Initial Steps; Switching On and Köhler Illumination

Leica DM2000
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
7. Start-up
7. Start-up
7.1 Switching on the Microscope
•Switch on the microscope with the on/off
switch (32.1, 33.5).
Caution!
After turning on the gas discharge lamp, the
burner must be immediately adjusted. There-
fore, do not turn on the power supply unit
yet. First, work in transmitted light in order to
familiarize yourself with the microscope’s
controls.
7.2 Köhler Illumination
The condenser is also pre-adjusted in the
factory.
However, in some cases it may be necessary to
re-adjust the condenser. Therefore, check the
condenser centering.
The following procedure is provided for
transmitted light-bright field illumination.
If present: click the condenser disk* into the
BF position.
If present: pull the light ring slide* out of the
condenser.
Select an objective with moderate
magnification (10x-20x).
For condensers with movable condenser
heads:
Swing in the condenser top.
(The condenser top is swung out for objective
magnifications < 10x.)
Insert the specimen in the stage’s specimen
holder.
Focus on the specimen using the focus wheel
(32.2).
21 3
Fig. 32 Leica DM2000
1 On/Off switch
2 Focus wheel
3 Stage positioning

Table of Contents