Leica DM2000 - Microscope Measurements: Gout Identification; Gout and Pseudo Gout Crystal Identification

Leica DM2000
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
59
10. Measurements with the Microscope
•Move the orientation handle to its extreme
right position. Now the N/S crystals should be
blue, and E/W yellow (fig. 68).
Be sure to test crystals with the orientation
handle in each position to insure positive
identification.
The following is the procedure for identifi-
cation of pseudo gout:
The test for pseudo gout is done identically to
the test for gout. However, the color change is
opposite that of Gout. That is, with the handle at
the left extreme, N/S crystals are blue and E/W
crystals are yellow, and vice versa with the level
at the right side (fig. 69).
Fig. 68 Identification of gout
Fig. 69 Identification of pseudo gout
Test for Gout
N/S crystals
yellow
E/W crystals
blue
Handle
left
N/S crystals
blue
E/W crystals
yellow
Handle
right
Test for Pseudo Gout
N/S crystals
blue
E/W crystals
yellow
Handle
left
N/S crystals
yellow
E/W crystals
blue
Handle
right