1.7 Acoustic signalisation
SOUND MEANING
CONTINUOUS SOUND overheating, scales overload, actuator overload, accumulator
overcurrent
REPEATED DISCONTINUOUS SOUND
(0,6s sound / 2,6s silence) STOP error (all STOP buttons are disabled)
(melody: 3 beeps, pause, 2 beeps, longer pause, 3 beeps,
pause, 2 beeps)
Bed Exit Alarm
DISCONTINUOUS SOUND
(duration 0,3s) conrmation
signalisation of stopping or signalisation of locked function
transition from tilt (Trendelenburg, Antitrendelenburg) to hori-
zontal position
Bed Exit Monitoring: transition from OFF to ON, transition from
PAUSE to ON
(duration 0,5s) start of service mode or end of service mode
keyboard error (positioning blocked)
(duration 3s) system error
(duration 3 min: 1,1s sound / 1,1s silence) Brake Signal
1.8 Illumination
NIGHT LIGHT
Bed illumination helps the nursing staff as well as the patient to orientate. The lowered intensity of lightning is set up after turning
the bed on. The night light is turned off during accumulator operation.
The bed is equipped with three-phase illumination:
1. Lowered intensity of illumination
2. Full intensity of illumination
3. Illumination is turned off
After pressing any button:
► The bed illumination will light up at full intensity for 10 minutes.
After 10 minutes the bed illumination will be lowered.
After disconnection of the bed from the mains illumination lights up for few seconds.
Turning off bed illumination:
► Disconnect bed from the mains.
After disconnection of the bed from the mains illumination lights up for few seconds.