D9U001GE4-0101_0520
Table 1 - IMMUNITY to RF wireless communications equipment
Test frequency (MHz) Band (MHz) Service Modulation Immunity Test Level
V/m
385 380 - 390 TETRA 400 Pulse modulation 18 Hz 27
450 430 - 470 GMRS 460, FRS 460 FM ± 5 kHz deviation 1
kHz sine
28
710
745
780
704 - 787 LTE band 13, 17 Pulse modulation 217
Hz
9
810
870
930
800 - 960 GSM 800/900, TETRA
800, iDEN 820, CDMA
850,
LTE band 5
Pulse modulation 18 Hz 28
1 720
1 845
1 970
1 700 - 1 990 GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE band 1, 3,
4, 25; UMTS
Pulse modulation 217
Hz
28
2 450 2 400 - 2 570 Bluetooth, WLAN,
802.11 b/g/n, RFID
2450,
LTE band 7
Pulse modulation 217
Hz
28
5 240
5 500
5 785
5 100 - 5 800 WLAN 802.11 a/n Pulse modulation 217
Hz
9
NOTE There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE There are no known other measures for keeping the basic safety based on EMC phenomena.
NOTE Beds equipped with communication module meet standard for IEEE 802.11 b/g/n (2400,0 MHz – 2483,5 MHz, modulation
DSSS (IEEE 802.11 b ), OFDM (IEEE 802.11 g/n) 20MHz bandwidth, EIRP = 0,34 W).
6 Use and Storage Conditions
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
► Ground the mains.
► Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
Eleganza 4 is designed for use in rooms for medical purposes. Electrical installations must therefore meet local norms laying down
the necessary conditions for electrical installations.
► Disconnect the bed from the mains in exceptional cases (i.e. lightnings, earthquake).
Eleganza 4 is not suitable for indoor environments containing ammable gases (except oxygen cylinders).