EasyManuals Logo
Home>LINET>Medical Equipment>WS17

LINET WS17 User Manual

Default Icon
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
D9U001GE4-0101_0526
9 Power Cable
Attachment plug is means of connecting and disconnecting bed from the mains.
Mains power cable must be attached with a hook at the head end of the bed during transport.
CAUTION!
Disconnecting bed from the mains does not stop motions of the bed!
Stop the bed before disconnection bed from the mains.
Where the integrity of the external protective conductor in the installation or its arrangement is in doubt
operate the bed from internal accumulator only.
10 Accumulator
WARNING!
When the bed is not connected to the mains and accumulator is not sufciently charged all electrical functions 
of the bed are blocked!
Purpose
The accumulator serves as a backup during power failures or for emergency bed positioning.
Use only accumulator approved by the manufacturer.
Check the functionality of the accumulator at least once a month in accordance with the user manual and have the
accumulator changed if necessary.
NOTE The service life of the accumulator depends on the frequency and method of use (2 years at the maximum).
The manufacturer will assume no responsibility for any damage to the bed or the accumulator caused by:
non-observance of the manufacturer’s instructions in the user manual
using accumulator not approved by the manufacturer
Warranty
The manufacturer provides a 6-month warranty for the full function of the accumulator.
Charging
The accumulator supplied with the bed is delivered insufciently charged.
Charging of the accumulator before use of the bed takes approximately 4 hours.
Charging of the accumulator runs automatically when the power cable is connected to the mains.
To charge the accumulator:
Connect the bed to the mains.
Storage
For declared lifetime period of leaded accumulator is recommended during storage:
To prevent accumulator from deep discharging and to keep accumulator at least partly charged by regular recharging
To store accumulator on the places with temperature from 10°C to 40°C
To prevent accumulator from being in the sunshine