D9U001GE4-0101_0582
18 Troubleshooting
DANGER!
Risk of mortal injury due to electric shock!
► If a fault occurs, have the electric motor, power box or other electrical parts repaired by qualied personnel
exclusively.
► Do not open the protective covers of the electric motor or the power box.
Error/Fault Cause Solution
Adjusting with position buttons not
possible
GO Button was not pressed Press the GO button.
Function disabled on Attendant Control Panel Enable disabled function.
Actuators have no power
Defective actuators
Defective accumulator
Check the mains connection.
Notify the service department.
Mains Plug inserted incorrectly Insert the Mains Plug correctly.
Faulty Power Source Notify the service department.
Faulty Control Element Notify the service department.
Faulty Mattress support platform Height/Tilt
Adjustment
Obstacle on the undercarriage cover Remove the obstacle.
Function disabled on Attendant Control Panel Enable disabled function.
Actuators have no power
Defective actuators
Defective accumulator
Check the mains connection.
Notify the service department.
Mains Plug inserted incorrectly Insert the Mains Plug correctly.
Faulty Power Source. Notify the service department.
Faulty Control Element. Notify the service department.
Lowering Backrest from the upright position
not possible
Obstacle under the Backrest or in the drive
mechanism.
Remove the obstacle
CPR Release Handle is defective Notify the service department.
Adjusting Siderails not possible Obstacle in the Siderail Release Mechanism Remove the obstacle.
Siderail Release Mechanism is defective. Notify the service department.
Faulty brakes Obstacle blocking brakes mechanically Remove the obstacle.
The brake mechanism is defective Notify the service department