11.4 Control Elements
The bed is operated by different control elements.
Control elements depending on the model:
■ Nurse Control Panel in both head siderails
■ iBoard Basic in both head siderails
■ Attendant Control Panel
■ Handset with illuminated buttons and with adapter for easy connection (Plug and Play)
■ Patient Control Panel in both foot siderails
■ Foot Switch - Bed Height
Disabling individual functions on the Attendant Control Panel or on the Nurse Control Panel will affect all control ele-
ments.
If the bed does not react to individual position settings:
► Check whether the function is disabled on the Attendant Control Panel or on the Nurse Control Panel.
POSITIONING Nurse
Control
Panel
Attendant
Control
Panel
Handset Patient Cont-
rol Panel
Foot Switch-
Bed Height
Backrest
a a a a
Thighrest
a a a a
Bed Height
a a a a
Autocontour
a a a a
Examination Position
a a
Emergency Trendelenburg
Position
a a
Anti-Trendelenburg and
Trendelenburg Tilt
a a
CPR Position
a a
Cardiac Chair Position
a a
Mobilization Position
a