SCPC System • Safety
8 CP_SEM_60-03-00.002
2.2.3 Safety instructions related to service
Note:
For additional information on general safety instructions as well as safety instructions related to use
and operation, refer to chap. 2 in the Operating Instructions.
Occupational health and safety
Correct handling of objects reduces mechanical stress to the spine and back muscles.
Do not wear metal jewellery such as rings, bracelets, etc., when performing maintenance and
service tasks.
Wear protective gloves when cleaning and disinfecting the SCPC System.
!
The SCPC System must not be used in the presence of explosive substances.
Instructions on the system
!
The SCPC System must only be used with the Stöckert centrifugal pump system (abbreviation:
SCP System).
The serial port is only intended to transmit DMS data (e.g. to another computer).
Make sure that the air circulation at the ventilation openings is unimpeded. Insufficient ventilation
may cause overheating.
Keep the following in mind when operating the SCPC system in combination with an SCP system
with a Revolution
®
and an Electrical Remote-Controlled Tubing Clamp:
If the speed exceeds 3200 RPM, the batteries are not charged in order to guarantee the power
supply to the Electrical Remote-Controlled Tubing Clamp.