SCPC System • Safety
8 CP_SEM_60-03-00.002
2.2.3 Safety instructions related to service
Note:
For additional information on general safety instructions as well as safety instructions related to use
and operation, refer to chap. 2 in the Operating Instructions.
Occupational health and safety
◗ Correct handling of objects reduces mechanical stress to the spine and back muscles.
◗ Do not wear metal jewellery such as rings, bracelets, etc., when performing maintenance and
service tasks.
◗ Wear protective gloves when cleaning and disinfecting the SCPC System.
!
◗ The SCPC System must not be used in the presence of explosive substances.
Instructions on the system
!
◗ The SCPC System must only be used with the Stöckert centrifugal pump system (abbreviation:
SCP System).
◗ The serial port is only intended to transmit DMS data (e.g. to another computer).
◗ Make sure that the air circulation at the ventilation openings is unimpeded. Insufficient ventilation
may cause overheating.
◗ Keep the following in mind when operating the SCPC system in combination with an SCP system
with a Revolution
®
and an Electrical Remote-Controlled Tubing Clamp:
If the speed exceeds 3200 RPM, the batteries are not charged in order to guarantee the power
supply to the Electrical Remote-Controlled Tubing Clamp.