PyroTec™ PRO Hardw are User’s Manual, V1.2 102
RT-MN024-1.2 02/21
Disposable tip tray empty
Wrong disposable tip tray
elected
Place disposable tip tray on speci-
fied position
Use adjust DiTi position function in
PyroTec™ PRO software module
See Application Software Manual
Wrong tip position coordi-
nates
Define tip position
See Application software manual
DiTi adapter mounted in-
correctly
Check correct installation of DiTi
adapter
See “Disposable Tip Cone (DiTi
Bad ground connection of
carrier
Place rack correctly on carrier
Clean carrier, to ensure good con-
nection
Assure container-rack-carrier-work-
ters
Check parameter in the application
software
See “Instrument Parts and Cleaning
MultiSense option: Pres-
sure measurement bad
Contact Lonza Scientific Support
Not enough liquid detected
Incorrect container/rack
Check container, rack definition
See Application Software Manual
Wrong container diameter
Check container data
See “Application Software Manual”
Contact Lonza Scientific Support
high level of static electric-
Do not use mobile phones, not even
in standby-mode closer than 2 m to
Increase ambient humidity (humidi-
fier)
Incorrect sample position-
ing
Bent tips
Use of incorrect carrier
Incorrect LiHa, X-, Y- and
Z-setup
Incorrect tip configuration
Rectify sample positioning
Replace bent tips
Use/configure the correct carrier.
Contact Lonza Scientific Support
Rectify tip configuration