PyroTec™ PRO Hardw are User’s Manual, V1.2 76
RT-MN024-1.2 02/21
ATTENTION
After a harsh crash some components of the instrument may be out of alignment or even
defective.
If a harsh crash occurred, contact your local service organization to have the instrument
checked.
RoMa Crash
After a crash with the RoMa, check the gripper and the RoMa alignment. Refer to 10.2.2
RoMa / Gripper Alignment”.
8.4 Operating in Process Definition Mode
8.4.1 Process Validation
The device must be validated by qualified technical personnel (e.g. Lonza approved engi-
neer or technician) before putting into service and after changes. Use of kits or kit compo-
nents on PyroTec™ PRO is only allowed after validation by Lonza, the kit manufacturer or
operator of the system.
For all applications of PyroTec PRO, the user must ensure that the requirements of each
protocol are carefully observed.
A systematic approach of risk analysis, validation of critical parameters and system valida-
tion should be followed to ensure that the system or combination with kit provides relia-
ble and reproducible performance.
Make sure that the validation process is executed according to national laws and stand-
ards.
ATTENTION
Crash or erroneous process results
The PyroTec™ PRO requires accurate positioning of all reagents, samples, racks, and plates
on the worktable.
Verify these positions accordingly before executing any program.
8.4.2 Liquid Handling
8.4.2.1 General Instructions
System Liquid
Make sure that the requirements for the system liquid are met.