Factory and Field Verification Values
Conditions
General test conditions, carried out with the Setup & Service Software:
• All liquid handling values have been verified under controlled laboratory environ-
ment at temperatures between 20°C and 27°C and a humidity between 30 % and
60 % at 25°C.
• Liquid: water
• Pipetting mode: single pipetting, free dispense
• New DiTi for each sample
The pipetting precision values listed in the table above are based on the following criteria:
• Liquid handling precision data tested for each produced PyroTec™ PRO
• Values are for pipetting water with disposable tips
• Default liquid class in EVOware
No single channel calibration required
Materials
In normal operation, the disposable tips are the only wetted parts. Other parts are not ex-
posed to the pipetted liquids except in case of a malfunction (too much liquid aspirated).
However, other parts may be exposed to aerosols from the liquid.
• The parts that come into contact with sample liquid:
o Disposable tips: PP
• The parts that may be moistened with aerosols:
o Tip cone (gold-plated brass)
4.4.2 Robotic Manipulator Arm (RoMa)
The PyroTec™ PRO instrument is equipped with a robotic manipulator arm. The robotic
manipulator arm is used to transport microplates from one position to another position on
the worktable.
Total range: 259 mm (10.2 in.)
Work range: 257 mm (10.12 in.)
58 to 140 mm (2.28 to 5.51 in.)
270° (left or right oriented)