PyroTec™ PRO Hardw are User’s Manual, V1.2 94
RT-MN024-1.2 02/21
Waste Bag
Specification
Typical dimensions for the waste bag (W x L) 300 mm x 600 mm
Thickness: 0.05 mm
Material: Polypropylene, polyethylene, or co-polymer (auto-clavable)
Imprint: Biohazard
Note: The waste bags used must meet your local safety guidelines.
9.3.3 DiTi Waste Station Unit
WARNING
Potentially infectious
Instrument parts may be contaminated with potentially infectious materials.
Follow basic biohazard precautions
Wear appropriate personal protective equipment, such as gloves, lab coats and pro-
tective eyewear.
9.3.3.1 Cleaning the DiTi Waste Slide
The discarded DiTis contain residues of sample and reagents, which contaminate the DiTi
waste slide.
Note: Heavy contamination of the slide might cause the DiTis to be stuck in the DiTi waste
slide.
Quick Cleaning
To clean the DiTi waste slide, proceed as follows:
Open the front safety panel.
Removing DiTi waste slide inset
Remove the cover (A) from the DiTi waste slide.
Remove the DiTi waste slide inset (B) from its hold.
Note: Hold a tissue under the DiTi waste slide inset when carrying it away to pre-
vent contaminated substances and DiTis from dropping to the floor.