PyroTec™ PRO Hardw are User’s Manual, V1.2 95
RT-MN024-1.2 02/21
DiTi waste slide
Spray some disinfectant on the inner
surface of the DiTi waste slide as
shown in the figure.
Suitable agents:
Check if there are residues of contami-
nation on the inner surface of the DiTi
waste slide.
If yes, schedule a thorough cleaning of
the parts as described below.
Reinstall the DiTi waste slide inset.
Make sure that the positioning pin of the DiTi waste slide is properly positioned in the
slot (D).
Reinstall the cover
Thorough Cleaning
To thoroughly clean the DiTi waste slide, perform the following procedure.
Remove the cover from the DiTi waste slide as described above.
Remove the DiTi waste slide inset from its holder as described above.
Put the DiTi waste slide inset and the cover into a basin filled with cleaning agent and
allow soaking for 30 minutes to 4 hours (depending on agent).
Allow the parts to dry.
Reinstall the DiTi waste slide inset as described above.
Reinstall the cover as described above.
9.3.3.2 Cleaning the Complete DiTi Waste Station Unit
The DiTi waste station unit may be contaminated with residues from reagents and sam-
ples, which must be removed.
Apart from the normal position (work position) the DiTi waste station unit can take the fol-
lowing positions:
If pulled to the mechanical stop: The front access panel can be opened, but the unit
cannot be removed.
If pulled to the middle position: The unit can be removed, but the front access panel
cannot be opened.