PyroTec™ PRO Hardw are User’s Manual, V1.2 95
RT-MN024-1.2 02/21
Spray some disinfectant on the inner
surface of the DiTi waste slide as
shown in the figure.
Suitable agents:
Check if there are residues of contami-
nation on the inner surface of the DiTi
waste slide.
If yes, schedule a thorough cleaning of
the parts as described below.
• Reinstall the DiTi waste slide inset.
• Make sure that the positioning pin of the DiTi waste slide is properly positioned in the
slot (D).
• Reinstall the cover
Thorough Cleaning
To thoroughly clean the DiTi waste slide, perform the following procedure.
• Remove the cover from the DiTi waste slide as described above.
• Remove the DiTi waste slide inset from its holder as described above.
• Put the DiTi waste slide inset and the cover into a basin filled with cleaning agent and
allow soaking for 30 minutes to 4 hours (depending on agent).
• Allow the parts to dry.
• Reinstall the DiTi waste slide inset as described above.
• Reinstall the cover as described above.
9.3.3.2 Cleaning the Complete DiTi Waste Station Unit
The DiTi waste station unit may be contaminated with residues from reagents and sam-
ples, which must be removed.
Apart from the normal position (work position) the DiTi waste station unit can take the fol-
lowing positions:
• If pulled to the mechanical stop: The front access panel can be opened, but the unit
cannot be removed.
• If pulled to the middle position: The unit can be removed, but the front access panel
cannot be opened.