9.4 Precision and Function Tests
9.4.1 Liquid Handling Performance Verification Testing
Lonza recommends the use of the QC kit for liquid handling performance verification test-
ing at least once per year during Preventative Maintenance by a Lonza approved Field Ap-
plication Specialist; a higher frequency may apply, based on the Quality & Regulatory re-
quirements of the laboratory using the PyroTec™ PRO pl atf orm.
The QC Kit is based on Artel’s patented Ratiometric™ Photometric technology, and is being
made available for Lonza’s customers.
Customer benefits include:
• Traceability of test results to the international standards.
• Robustness of the method in a daily lab environment
• Ease-of-use
Methods
Gravimetric test; contact Scientific Support for more information on this method.
for Various
Tests
The “Setup & Service” and application software offer a variety of tests to check individual
functions of the modules.
Refer to the “Instrument Software Manual”
Refer to the “PyroTec™ PRO Software User Guide”
A Arm guide
B Arm guide roller
C
Arm rail
D Arm