PyroTec™ PRO Hardw are User’s Manual, V1.2 97
RT-MN024-1.2 02/21
• Wipe the surface of the worktable with a suitable cleaning (e.g. alcohol, disinfectant)
to remove any spilled reagent.
• If necessary, clean with additional water.
9.3.5 Safety Panels
Safety Panels
Perform the following procedure to clean the safety panels.
• Wipe the inner and outer surface of the safety panels with a suitable cleaning agent,
e.g. water, alcohol, or disinfectant, to remove any spilled reagent or sample.
• If necessary, further clean the surface with water or alcohol.
9.3.6 Lower DiTi Eject Option
Rocker
Perform the following procedure to clean the rocker of the lower DiTi eject option:
• Wipe the surface of the rocker (A) with a suitable cleaning agent (e.g. alcohol, disin-
fectant)
• If necessary, further clean with water.
9.3.7 Arm Guide
The following description is applicable to:
• Liquid handling arm (LiHa)
• Robotic manipulator arm (RoMa standard and long)
Arm Guide
In order to avoid uneven movements of the arm, use a cotton tab or a lint free tissue on a
screwdriver to clean the arm guide roller and a lint free tissue to clean the arm rails.
Note: Do not use alcohol or solvents to clean the arm guide. Do not use grease on the arm