PyroTec™ PRO Hardw are User’s Manual, V1.2 22
RT-MN024-1.2 02/21
Damaged, lost, or illegible symbols (notices or stickers) must be replaced immediately.
3.3.2 Laser Radiation
The optional barcode scanner emits laser radiation—namely, a low-power collimated
beam in the visible spectrum with the following properties:
• Wavelength: 655 nm
• Pulse duration: 150 µs
• Maximum power of energy output: 1.0 mW
The Loading-ID is a class 1 laser product pursuant to IEC/EN 60825‑1:2014 that emits laser
radiation.
Dazzle, flash-blindness, and after images may be caused by the laser beam. Skin or iris in-
jury can be caused near to or in contact with the source.
• Do not stare into the laser beam or into its reflections.
• Do not reach into the laser beam near to the source.
•
Do not touch the source during operation.
3.4 Decontamination Declaration
contaminate
Apart from regular decontamination, the user must thoroughly decontaminate the instru-
ment according to standard laboratory regulations in the following cases:
• Before any maintenance or service work is performed on the instrument.
• In case of accidents (e.g. crash, spilled substances, etc.).
• Before a Lonza approved FAS (field application specialist) performs any on-site work
on the instrument.
• Before the instrument or parts of it are returned to Lonza (e.g. for repair).
• Prior to storage of the instrument.
• Prior to disposal of the instrument or parts of it.
• Generally, before the instrument or parts of it leave the user’s site.
tion Method
The decontamination method must be adapted to the respective application and the sub-
stances associated with it. The user takes the full responsibility for the appropriate decon-
tamination of the entire equipment.