PyroTec™ PRO Hardw are User’s Manual, V1.2 49
RT-MN024-1.2 02/21
Note: though the instrument works perfectly at this minimum distance to the wall, be
aware of the fact that it is an advantage if the rear of the instrument is accessible, e.g. for
maintenance and service work.
Weights
Load capaci-
ties
Check the minimal load capacities of:
Floor
365 kg / 804.7 lbs.
Table or cabinet
300 kg / 661.4 lbs.
Which
components
take wall
outlets
Supply
Check the availability of wall outlets
Instrument
Computer
Readers
Workbench
ATTENTION
PyroTecPRO placed on a weak and unstable workbench can lead to LLD and pipetting
failures caused by vibrations of the instrument or other devices on the workbench.
Place the instrument on a rigid vibration-free workbench that is able to carry the
weight of the instrument.
Make sure that the surface of the workbench does not get out of shape due to the
load. All feet of the instrument must rest firmly on the surface for levelling the instru-
ment properly.
Environmental Conditions
Make sure that the environmental conditions, such as humidity, operating temperature,
etc. are fulfilled.
If the instrument is equipped with a LoadingID, make sure that the intensity of external
light does not exceed 8000 lux.
ATTENTION
Barcodes cannot be read in the presence of sunlight or other strong light sources.
Do not install the instrument where it is exposed to direct sunlight.
Strong light sources may impair the function of the barcode scanner.