PyroTec™ PRO Hardw are User’s Manual, V1.2 91
RT-MN024-1.2 02/21
o Replace the inline filter.
Settings
• To ensure operating readiness, perform the following test:
o Instrument Warm-up
Refer to the PyroTec™ PRO Software User Guide.
Please contact your local Lonza representative to schedule a service visit.
9.3.1.3 Air LiHa Inline Filter
9.3.1.4
Note: Irrespective of the regular maintenance schedule, the filter needs replaced if moi s-
tened because of an erroneous aspiration (too much liquid aspirated).
DiTi cone can be contaminated.
• Decontaminate the DiTi cone and assure appropriate safety measures.
•
The filter removal tool can cause injuries.
• Avoid piercing your finger or hand.
• Wear protective rubber gloves during filter removal.
To replace the inline filter in the DiTi cone of the Air LiHa, proceed as follows:
• Remove the inline filter from the DiTi cone as shown in the figure below.