Lumibird Quantel ABSOLU
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service manual
Service Manual 30 / 63
XE_ABS_MT_AN_200224
10.4. Check the image of the rings (only B20-5A)
1. Delete the B20-5A probe parameters installed on the ABSolu Software: right click on
the line and select “delete”:
2. Install the “B20_Serv.uprb” file you will find under the following link:
“C:\Quantel\ABSolu\AppData\ConfigFiles\ProbesFiles”: