Position
Converter Function Code Function Name Function Description
Access
Level
MCP
Command
6.6
[Meas.lock=OFF]
[Meas.lock=OFF] When this function is active (ON),
applying a voltage to the on
input terminals, the measurement
is stopped and the meter will
display zero ow.
AL3 [MSLIE]
6.7
[Calibration=OFF]
[Calibration=OFF] When this function is active,
applying a voltage to the on/
o input terminals, the meter
performs an autozero calibration
cycle.
AL3 [CALIE]
Note: If the voltage pulse is less than 1 second, the meter performs a calibration cycle to compensate
for possible thermal drifts. If the voltage pulse is more 1 second, the meter performs a zero calibration
measure. To perform the calibration, it is absolutely necessary for the pipe to be full of liquid and that the
liquid is perfectly still. Even very small movement of the liquid may aect the result of the calibration and
consequently the accuracy of the system.
6.8
[Range
change=OFF]
[Range
change=OFF]
Enables the range change
external command. When this
function is enabled, applying
a voltage to the on/o input
terminals, the meter switches to
the second measuring range (Fs2).
AL3 [SRCIE]
Note: The autorange does not allow manually changing the range. See POS. 4.4.