EasyManua.ls Logo

McCrometer ProComm - 8 Alarm Messages

McCrometer ProComm
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.0 ALARM MESSAGES
MESSAGE CAUSES ACTION TO TAKE
NO ALARMS All works regularly ---
[000] SYSTEM RESTART --- ---
[001] INTERNAL PS FAIL Internal supply voltage error Contact the service
[002] CLOCK NOT SET System Clock not set Set the system clock from the converter
menu 13 (see also MCP function ).
[003] SD CARD FAILURE SD card not found or unreadable Check and/or replace SD card
[005] F-RAM ERROR Error writing / reading Flash-RAM Contact customer service
[006] EXCITATION ERROR The excitation of the sensor coils resulting
from cable is interrupted
Check the connecting cables to the sensor
[007] SIGNAL ERROR The measure is strongly eected by
external noise or the cable connecting the
converter to the sensor is broken.
Check the status of the cables connecting
the sensor and the device grounding
connectionsfor possible presence of noise
sources.
[008] PIPE EMPTY The measuring pipe is empty or the
detection system has not been properly
calibrated.
Check whether the pipe is empty, or repeat
the empty pipe calibration procedure.
[009] FLOW>MAX+ The ow rate is higher than the maximum
positive threshold set.
Check the maximum positive ow rate
threshold set and the process conditions.
[010] FLOW>MAX- The ow rate is higher than the maximum
negative threshold set.
Check the maximum negative ow rate
threshold set and the process conditions.
[011] FLOW<MIN+ The ow rate is lower than the minimum
positive threshold set.
Check the minimum positive ow rate
threshold set and the process conditions.
[012] FLOW<MIN- The ow rate is lower than the minimum
negative threshold set.
Check the minimum negative ow rate
threshold set and the process conditions.
[013] FLOW>FULL SCALE+ The ow rate is higher than the full scale
positive value set on the instrument.
Check the full scale positive value set on the
instrument and the process conditions.
[014] FLOW>FULL SCALE- The ow rate is higher than the full scale
negative value set on the instrument.
Check the full scale negative value set on
the instrument and the process conditions.
[015] PULSE1>RANGE The pulse generation output 1 of the
device is saturated and cannot generate
the sucient number of impulses.
Set a higher unit of volume or, if the
connected counting device allows it, reduce
the pulse duration value.
[016] PULSE2>RANGE The pulse generation output 2 of the
device is saturated and cannot generate
the sucient number of impulses.
Set a higher unit of volume or, if the
connected counting device allows it, reduce
the pulse duration value.
[017] CALIBR.ERROR Calibration Error Contact customer service
[018] SYSTEM FREQ.ERR System Freq. Error Contact customer service
[019] B.DATA NOT INIT Uninitialized data system Contact customer service
[020] FL.SENSOR ERROR Flow rate sensor error Contact customer service
[021] BATTERY LOW (Rechargeable) battery depleted Contact customer service to replace the
battery
30124-60 Rev. 1.0 | unreleased
Page 40
ALARM MESSAGES
ProComm
Preliminary

Other manuals for McCrometer ProComm

Related product manuals